MURAGAKI Yoshihiro
   Department   School of Medicine(Tokyo Women's Medical University Hospital), School of Medicine
   Position   Visiting Professor
Language Japanese
Title Development of the Hybrid Marker System for Auto Registration in the
MRI-guided Surgical Navigation (Second study on clarification of adequate
space for setting the hybrid markers)
Conference Type Nationwide Conferences
Presentation Type Speech
Date 2008/10/31
Society abstract 日本コンピュータ外科学会誌 10(3),401-402 2008
Summary Abstract: For pre-clinical study of the hybrid marker, we fabricated a phantom made of an acrylic resin and a
contrast agent for MRI. Two hundred holes were drilled in the acrylic resin plate. Distance between two holes
was 20 mm. Each hole was filled with 170 micro litters of the contrast agent. We took MRI images of phantom
for clarification of adequate space for setting the hybrid markers. Imaging sequence was the same as that
during neurosurgery. Slice thickness was 1 mm. As a result, image distortion was observed when the hole was
located at more than 40 mm far from the center of MRI gantry. It was shown that measured error was less than
1mm in adequate space. The space was located 40 mm above from the center of MRI gantry, and 120 mm from
the receiver coil for MRI.