ムラガキ ヨシヒロ
Muragaki Yoshihiro
村垣 善浩 所属 医学部 医学科(東京女子医科大学病院) 職種 客員教授 |
|
言語種別 | 英語 |
発表タイトル | A surgical navigation system for aortic vascular surgery: A practical approach |
会議名 | 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society |
発表形式 | 口頭 |
発表者・共同発表者 | ◎UEMATSU Miyuki, ICHIHASHI Takuya, ASATO Kenya, NAKAOKA Ryusuke, MATSUOKA Atsuko, AOMI Shigeyuki, IIMURA Hiroshi, SUZUKI Takashi, MURAGAKI Yoshihiro, ISEKI Hiroshi |
発表年月日 | 2013/07 |
開催地 (都市, 国名) |
Osaka, Japan |
学会抄録 | Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE 5327-5330 |
概要 | Abstract
In aortic vascular surgery, a navigation system must represent the anatomical map of individual patient in order to detect the important artery. To provide a proper fit for positions along the dorsoventral axis, the spinous process was added to a currently used anatomical point set consisting of four anterior body landmarks. In addition, we attempted to reduce the registration error by compensating for alignment errors resulting from variations in tissue thickness at each landmark. The alignment values were examined using a human phantom consisting of a skeleton model with subcutaneous tissue in the semilateral position. Using this method, a phantom simulation and five clinical trials were performed. Target errors were evaluated at the orifice of the intercostal artery. In the phantom simulation, the error at the target point was 4.1 ± 2.7 mm. However, for one patient undergoing thoracoabdominal aortic aneurysm replacement surgery, the target error was 8.0 mm using the proposed method. |