ムラガキ ヨシヒロ
Muragaki Yoshihiro
村垣 善浩 所属 医学部 医学科(東京女子医科大学病院) 職種 客員教授 |
|
言語種別 | 英語 |
発表タイトル | Robotic tensor device for optimal soft tissue balance in TKA. |
会議名 | 35th Annual International Conference of the IEEE Engineering in Medicine and Biology Society. |
発表形式 | 口頭 |
発表者・共同発表者 | ◎OKAMOTO Jun, MATSUMOTO Tomoyuki , MURATSU Hirotsugu, KUROSAKA Masahiro, MURAGAKI Yoshihiro, ISEKI Hiroshi |
発表年月日 | 2013/07 |
開催地 (都市, 国名) |
Osaka, Japan |
学会抄録 | Engineering in Medicine and Biology Society (EMBC), 2013 35th Annual International Conference of the IEEE 6695-6698 |
概要 | Abstract
Soft tissue balance measurement is necessary for effective total knee arthroplasty (TKA). If optimal ligament balance is not achieved during surgery, it results in laxity of the knee or repeat surgery. In this paper, a newly developed robotic tensor system for measurement of soft tissue balance during TKA is described. The development concepts of the system were to measure collateral ligament balance (joint gap distance and varus angle) after cutting of the tibia and femur and to generate arbitrary lifting force to the knee joint during surgery. The system was created according to the development concepts. The prototype system has an active prismatic joint and a passive rotation joint to assess the joint gap and varus angle. The maximum lifting force is 250 N, the measurement range of the joint gap is 30 mm, and the varus angle is ± 15°. To confirm the force control ability, a basic experiment was performed. The input target force was 100, 150, 200, 250 N, and the output result showed an error within 3N. This experimental result shows that the newly developed system appears to have better accuracy than previous instrumental methods. |