イチバ シンゴ   Ichiba Shingo
  市場 晋吾
   所属   医学部 医学科(東京女子医科大学病院)
   職種   教授
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 Mean lung pressure during adult high-frequency oscillatory ventilation: an experimental study using a lung model.
掲載誌名 正式名:Acta medica Okayama
略  称:Acta Med Okayama
ISSNコード:0386300X/0386300X
掲載区分国内
巻・号・頁 68(6),pp.323-329
著者・共著者 Hirayama Takahiro, Nagano Osamu, Shiba Naoki, Yumoto Tetsuya, Sato Keiji, Terado Michihisa, Ugawa Toyomu, Ichiba Shingo, Ujike Yoshihito
発行年月 2014/12
概要 In adult high-frequency oscillatory ventilation (HFOV), stroke volume (SV) and mean lung pressure (PLung) are important for lung protection. We measured the airway pressure at the Y-piece and the lung pressure during HFOV using a lung model and HFOV ventilators for adults (R100 and 3100B). The lung model was made of a 20-liter, airtight rigid plastic container (adiabatic compliance: 19.3 ml/cmH2O) with or without a resistor (20 cmH2O/l/sec). The ventilator settings were as follows: mean airway pressure (MAP), 30 cmH2O; frequency, 5-15 Hz (every 1 Hz); airway pressure amplitude (AMP), maximum;and % of inspiratory time (IT), 50% for R100, 33% or 50% for 3100B. The measurements were also performed with an AMP of 2/3 or 1/3 maximum at 5, 10 and 15 Hz. The PLung and the measured MAP were not consistently identical to the setting MAP in either ventilator, and decreasing IT decreased the PLung in 3100B. In conclusion, we must pay attention to the possible discrepancy between the PLung and the setting MAP during adult HFOV.
DOI 10.18926/AMO/53021
PMID 25519026