イセキ ヒロシ   ISEKI Hiroshi
  伊関 洋
   所属   研究施設 研究施設
   職種   非常勤講師
論文種別 原著
言語種別 英語
査読の有無 査読なし
表題 New radiofrequency coil integrated with a stereotactic frame for intraoperative MRI-controlled stereotactically guided brain surgery
掲載誌名 正式名:Stereotactic and functional neurosurgery
略  称:Stereotact Funct Neurosurg
ISSNコード:10116125/14230372
巻・号・頁 84(4),pp.136-141
著者・共著者 TANIGUCHI Hiroki†, MURAGAKI Yoshihiro, ISEKI Hiroshi, NAKAMURA Ryoichi, Taira Takaomi, TAIRA Takaomi*
発行年月 2006/08/04
概要 Mislocalization errors caused by MR image distortions or brain shift are one of the main causes of complications after stereotactically guided neurosurgical procedures. A special device, which could ameliorate such effects and provide intraoperative acquisition of MR images of sufficient diagnostic accuracy, was developed. It is composed of a radiofrequency receiver coil integrated with a modified Komai stereotactic frame and a Sugita four-pin head holder. Clinical testing revealed that the use of the device during stereotactically guided procedures under the control of low magnetic field strength (0.3 T) intraoperative MRI ameliorates the effects of brain shift, and permits to obtain informative tissue samples and to perform aggressive removal of the lesion without any damage of the eloquent cerebral structures. The possibility of an easy shift from a stereotactically guided to a routine microneurosurgical procedure represents an additional benefit of the developed device.
DOI 10.1159/000094845
PMID 16899977