Title Development of a Semiautomatic Dura Mater Suturing Device for Preventing Cerebrospinal Fluid Leakage in Transsphenoidal Surgery.
Journal Formal name:Surgical innovation
Abbreviation:Surg Innov
ISSN code:15533514/15533506
Domestic / ForeginForegin
Volume, Issue, Page 28(3),pp.374-377
Author and coauthor SUZUKI Shohei†, KOBAYASHI Etsuko, HODOTSUKA Kenichi, AMANO Kosaku, MASAMUNE Ken*, MURAGAKI Yoshihiro
Authorship Corresponding author
Publication date 2020/11/25
Summary Dural suturing in transsphenoidal surgery requires well-honed technical skills. We have developed a semiautomatic dural suturing device and confirmed its effectiveness by comparing it with the conventional method. This device significantly shortens the suturing time compared with the conventional method. The dural suturing time in transsphenoidal surgery could be decreased significantly by use of this novel device.