イカリ カツノリ
  猪狩 勝則
   所属   医学部 医学科(東京女子医科大学病院)
   職種   寄附部門教授
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 Effect on patellar kinematics of the different patellar component designs in total knee arthroplasty: intraoperative measurement of dome type versus anatomic type.
掲載誌名 正式名:European journal of orthopaedic surgery & traumatology : orthopedie traumatologie
略  称:Eur J Orthop Surg Traumatol
ISSNコード:16338065/16338065
掲載区分国外
巻・号・頁 e-pub,pp.e-pub
著者・共著者 Mochizuki Takeshi, Yano Koichiro, Ikari Katsunori, Hiroshima Ryo, Okazaki Ken
発行年月 2019/10
概要 BACKGROUND:The aim of this study was to identify, during total knee arthroplasty surgery, the effect on patellar kinematics of different patellar component designs in the same patients.METHODS:This study enrolled 84 patients with osteoarthritis. Intraoperative X-rays were used to measure internal rotation angle, flexion angle and lateral tilt were at different knee flexion angles with dome-type or anatomic-type patellar components (ATTUNE®, DePuy).RESULTS:Significant differences from baseline between the two types of components occurred at 120° of the knee flexion in the internal rotation angle, at 90° and 120° of the knee flexion in the flexion angle, and at 60° of the knee flexion in the lateral tilt.CONCLUSION:This study revealed that the difference in patellar component design affects patellar kinematics. Therefore, the difference in patellar component design may affect patellar stability.
DOI 10.1007/s00590-019-02586-1
PMID 31655867