ニシザワ ヨウコ   Nishizawa Yoko
  西沢 蓉子
   所属   医学部 医学科(附属足立医療センター)
   職種   非常勤講師
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 Study of discrepancies between recorded and actual blood flow in hemodialysis patients.
掲載誌名 正式名:ASAIO journal (American Society for Artificial Internal Organs : 1992)
略  称:ASAIO J
ISSNコード:(1538-943X)1058-2916(Linking)
掲載区分国外
巻・号・頁 59(6),pp.617-21
著者・共著者 Kimata Naoki, Wakayama Koji, Okano Kazuhiro, Hibi Asako, Sawada Anri, Tajima Yoko, Makabe Shiho, Tsuchiya Ken, Mineshima Michio, Nitta Kosaku, Akiba Takashi
担当区分 責任著者
発行年月 2013/11
概要 Adequate blood flow (Qb) is necessary for effective hemodialysis (HD). Aim of the study was to examine relationship between the actually delivered Qb (dQb) and reported Qb (rQb) with dialysis machine. One hundred HD patients with arteriovenous fistula were enrolled. Delivered Qb was measured at the beginning and end of each HD session. dQb/rQb < 1 indicated a discrepancy between actual dQb and rQb reported using a dialysis machine. In addition, dQb/rQb was examined in HD patients using needles of different gauges during treatment. The average levels of dQb/rQb at start and end of HD session were 1.01 ± 0.04 and 0.98 ± 0.05, respectively. In the 16 gauge and 17 gauge needle groups, the percentage of patients with dQb/rQb < 1 increased in accordance with the increase in rQb or as the HD session progressed. In the 15 gauge needle group, the percentage of patients with dQb/rQb < 1 was <50% at any level of rQb. Selection of needle gauge is important factors for determining actual dQb in HD patients.
DOI 10.1097/MAT.0b013e3182a708b9
PMID 24172267