表題 | Spatial factors for quantifying constant flow velocity in a small tube phantom: comparison of phase-contrast cine-magnetic resonance imaging and the intraluminal Doppler guidewire method |
掲載誌名 | 正式名:Jpn J Radiol ISSNコード:1867-108X |
巻・号・頁 | 27(9),pp.335-41 |
著者・共著者 | Machida, H., Komori, Y., Ueno, E., Shen, Y., Hirata, M., Kojima, S., Morita, S., Sato, M., Okazaki, T. |
発行年月 | 2009 |
概要 | We examined the spatial factors influencing magnetic resonance (MR) flow velocity measurements in a small tube phantom and used the same measurements obtained with an intraluminal Doppler guidewire as reference.|We generated constant flow velocities from approximately 40 to 370 cm/s in a tube 4 mm in diameter. We then performed segmented k-space, phase-contrast cine-MR imaging to quantify spatial peak flow velocities of one pixel and of five adjacent pixels as well as spatial mean velocities within regions of interest in a cross section of the phantom. Pixel dimensions ranged from 1.00 x 1.00 mm to 2.50 x 2.50 mm. We compared the MR measurements with the temporally averaged Doppler spectral peak velocities.|For one pixel (r > 0.99: MR flow velocity for pixel dimension 1.00 x 1.00 mm = 1.03x + 9.8 cm/s), the linear correlation was excellent between flow velocities by MR and Doppler guidewire methods. However, for the five adjacent pixels, MR measurements were significantly underestimated using pixels 1.25 x 1.25 mm to 2.50 x 2.50 mm and for mean velocities for all pixel dimensions.|Relatively high spatial resolution allows accurate MR measurement of constant flow velocity in a small tube at spatial peak velocities for one pixel. |
DOI | 10.1007/s11604-009-0349-9 |
文献番号 | 19943143 |