イセキ ヒロシ
Iseki Hiroshi
伊関 洋 所属 医学研究科 医学研究科 (医学部医学科をご参照ください) 職種 特任顧問 |
|
論文種別 | 原著 |
言語種別 | 英語 |
査読の有無 | 査読なし |
表題 | New radiofrequency coil integrated with a stereotactic frame for intraoperative MRI-controlled stereotactically guided brain surgery |
掲載誌名 | 正式名:Stereotactic and functional neurosurgery 略 称:Stereotact Funct Neurosurg ISSNコード:10116125/14230372 |
巻・号・頁 | 84(4),pp.136-141 |
著者・共著者 | TANIGUCHI Hiroki†, MURAGAKI Yoshihiro, ISEKI Hiroshi, NAKAMURA Ryoichi, Taira Takaomi, TAIRA Takaomi* |
発行年月 | 2006/08/04 |
概要 | Mislocalization errors caused by MR image distortions or brain shift are one of the main causes of complications after stereotactically guided neurosurgical procedures. A special device, which could ameliorate such effects and provide intraoperative acquisition of MR images of sufficient diagnostic accuracy, was developed. It is composed of a radiofrequency receiver coil integrated with a modified Komai stereotactic frame and a Sugita four-pin head holder. Clinical testing revealed that the use of the device during stereotactically guided procedures under the control of low magnetic field strength (0.3 T) intraoperative MRI ameliorates the effects of brain shift, and permits to obtain informative tissue samples and to perform aggressive removal of the lesion without any damage of the eloquent cerebral structures. The possibility of an easy shift from a stereotactically guided to a routine microneurosurgical procedure represents an additional benefit of the developed device. |
DOI | 10.1159/000094845 |
PMID | 16899977 |