オカザキ ケン   KEN OKAZAKI
  岡崎 賢
   所属   医学部 医学科(東京女子医科大学病院)
   職種   教授・基幹分野長
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 O-Arm Assisted Cervicothoracic Spine Pedicle Screw Placement Accuracy Is Higher Than C-Arm Fluoroscopy.
掲載誌名 正式名:World neurosurgery
略  称:World Neurosurg
ISSNコード:18788769/18788750
掲載区分国外
巻・号・頁 158,pp.e996-e1001
著者・共著者 Inoue Tomohisa, Wada Keiji, Tominaga Ayako, Tamaki Ryo, Hirota Tomoya, Okazaki Ken
発行年月 2022/02
概要 OBJECTIVE:We compared the accuracy of C-arm fluoroscopy versus O-arm-assisted pedicle screw (PS) placement in the cervicothoracic spinal junction (CTSJ).METHODS:Patients who underwent PS placement in the CTSJ (C7-T4) at our hospital were included in this study. Of 37 patients who underwent PS placement in the CTSJ, 20 underwent intraoperative C-arm fluoroscopy-assisted surgery (C Group) and 17 underwent intraoperative O-arm-assisted surgery (O Group). In total, 159 PSs were placed-73 in the C Group and 86 in the O Group. The accuracy of PS placement was compared between the C Group and O Group using the classification proposed by Gertzbein and Robbins to analyze pedicle violation.RESULTS:PS accuracy was higher in the O Group than C Group; PS placement evaluated as grade A, representing no perforation, was 95.3% (82/86) for the O Group, whereas it was 78.1% (57/73) for the C Group. There was a clear statistically significant difference in accuracy of PS placement between the groups (P = 0.0013).CONCLUSIONS:O-arm-assisted surgery improved the accuracy of PS placement in the CTSJ.
DOI 10.1016/j.wneu.2021.12.006
PMID 34890848