モリタ サトル   Satoru Morita
  森田 賢
   所属   医学部 医学科(東京女子医科大学病院)
   職種   講師
論文種別 原著
言語種別 英語
査読の有無 査読あり
表題 Aliasing artifacts with the BLADE technique: causes and effective suppression
掲載誌名 正式名:J Magn Reson Imaging
ISSNコード:1522-2586
巻・号・頁 33(2),pp.432-40
著者・共著者 Kojima, S., Morita, S., Ueno, E., Hirata, M., Shinohara, H., Komori, A.
担当区分 2nd著者
発行年月 2011
概要 To elucidate the causes of aliasing artifacts with the BLADE technique and clarify the effective suppression methods.|We separately observed the aliasing artifacts of BLADE from features inside and then outside the defined field-of-view (FOV) using phantom studies. The effectiveness of suppressing them with phase oversampling (POS) and presaturation pulses (SAT) was evaluated. Finally, our observations were confirmed for a healthy volunteer.|Characteristic aliasing artifacts were observed from both inside and outside the FOV. Those from inside the FOV were sufficiently suppressed by using a POS of 25%, considering the acquisition time prolongation. Those from outside the FOV were nearly suppressed using SAT outside the FOV without selecting needless receiver coils. Aliasing artifacts on the coronal images of the head and neck with a healthy volunteer were completely suppressed by a combination of using a POS of 25% and using SAT on all four sides outside the FOV.|The characteristic aliasing artifacts of BLADE are caused from both inside and outside the defined FOV. They can be effectively suppressed by a combination of using a POS of 25% and using SAT on all four sides outside the FOV.
DOI 10.1002/jmri.22465
文献番号 21274986